молекулярні іони, молекулярные ионы ; аналітична хімія, аналитическая химия, analytical chemistry
The paper discusses influence of the molecular ions on the detection limit at determination of Rare Earth Elements and some toxic elements in mine waters of the Donetsk region using glow discharge mass-spectrometry (GD-MS). High resolution mass spectrometer was used to study impact of individual molecular ions formed from components of the sample and discharge gas Ar. Mass spectra were calculated in the area of masses of isotope studied and the most suitable isotopes were selected. It is shown that thesensitivity of analysis depends on the overall composition and isotope and is within 0.1- 20 ppb at resolving power of the instrument of the order of 5000. Simulation of mass-spectra allows
avoiding erroneous interpretation of mass spectra and correctlyestimate detection limit of analysis.
В роботі розглядається вплив молекулярних іонів на межу визначення РЗЕ та деяких токсичних елементів в шахтних водах Донецького регіону з використанням мас-спектрометрії жевріючого розряду. За допомогою спектрометрависокої роздільної здатності вивчався вплив окремих молекулярних іонів, які утворюються з компонентів зразка та розрядного газу аргону. Були розраховані мас-спектри в ділянках мас досліджуваних ізотопів та визначено найбільш придатні для аналізу ізотопи. Показано, що чутливість аналізу залежить від загального складу та ізо-топу і при роздільній здатності порядку 5000 знаходиться в межах 0.1 - 20 ppb. Моделювання мас-спектрів дозволяє уникнути помилкової інтерпретації мас-спектрів та правильно визначити межу визначення.